Determination of Volatile Organic Compounds (VOC) by Gas Chromatography/Mass Spectrometry (GC/MS)
The method for the determination of volatile organic compounds (VOC) by Gas Chromatography/ Mass Spectrometry (GC/MS) is for use in the analysis of samples regulated by Regulation XI.
1. Principle
2. Equipment
3. Reagents
4. Procedure
5. Calculations
1. Principle
The headspace or liquid portion of a sample is injected directly
into the gas chromatograph/mass spectrometer for quantitation
of individual compounds. The total VOC is the sum of the concentrations
of the individual compounds. A sample has VOC present if it contains
components above ambient levels. Ambient levels is defined as
less than 2 ppm total nonmethane hydrocarbons, as methane. [Return to Top]
2. Equipment
2.1 GC/MS equipped with capillary open split interface
2.2 Column: DB-624R, 30m X 0.53 mm, 3.0 um film
2.3 Syringes:
2.3.1 Glass, 1 mL with Luer tip syringe valve and 2-inch, 25
gauge needle
2.3.2 Gas-tight, 10 uL with needle
2.3.3 Gas-tight, 1.0 mL with needle
2.4 Volumetric Glassware
2.4.1 10 mL Volumetric Flask, Class A
2.4.2 1 mL Volumetric Pipets, Class A
2.5 Sample Containers
2.5.1 Jars, 250 mL, wide-mouth with TeflonR faced septa
2.5.2 Vials, 40 mL, screw cap with TeflonR faced septa
2.5.3 Cans, metal, 1 pint or 1 quart
2.6 Miscellaneous Supplies
2.6.1 Septa, ThermogreenTM, 11 mm
2.6.2 Glass injection port liners
2.6.3 Vials, 5 mL, screw cap with TeflonR faced septa
[Return to Top]
3. Reagents
3.1 Helium gas, chromatographic quality
3.2 VOC standards, >99% purity, as needed
3.3 Standard solutions, prepared by combining compounds with
similar functional groups.
3.3.1 Oxygenated organic compounds
3.3.2 Hydrocarbon compounds
3.3.3 Aromatic compounds
3.3.4 Chlorinated compounds
3.4 4-Bromofluorobenzene (BFB), 25 mg/L in methanol
[Return to Top]
4. Procedure
4.1 Set up the GC oven program for the GC/MS. For a Hewlett-Packard
5970B, the following oven program is recommended:
Injector Temp:
|
200°C
|
Transfer line temp:
|
280°C
|
Detector Temp:
|
250°C
|
Oven Temp:
|
Hold at 50°C for 5 min.
|
Ramp 15°C/min. to 200°C
|
Hold at 200°C for 15 min
|
Total Run Time: 30 min.
|
Scan start time:
|
0.0 min.
|
Scan Range:
|
5-250 amu
|
4.2 The system is checked for air leaks.
4.2.1 The system may be checked for air leaks either by running
the program without any injection or by looking at the amount
of air present during the optimization run using PFTBA. Air peak
ions should meet the following criteria:
m/z amu
|
Ion Abundance Criteria
|
Intensity (counts)
|
18
|
ion of Water (<1% of 69 amu of PFTBA)
|
<200
|
28
|
ion of Nitrogen (<1% of 69 amu of PFTBA)
|
<500
|
32
|
ion of Oxygen (<1% of 69 amu of PFTBA)
|
<200
|
4.2.1.1 When the intensities of the water, nitrogen and oxygen
ions are above their specified levels, it is an indication of
a leak in the system.
4.2.1.2 Apply Freon 22 to fittings and tighten until no 51
amu signal is detected.
4.3 Check the tune by using perfluorotributylamine (PFTBA) prior
to the start of an analysis or once a week if the instrument is
not used daily.
4.3.1 The PFTBA ions should meet the following:
m/z amu
|
Ion Abundance Criteria
|
Intensity (counts)
|
69
|
Base Peak 100 %
|
75,000-200,000
|
219
|
30 - 60 % of 69 amu
|
22,500-120,000
|
502
|
- 10 % of 69 amu
|
1750-20,000
|
4.3.1.1 The electron multiplier voltage is raised when the
69 amu intensity falls below 75,000 counts. Set the intensity
count to about 150,000 when adjustment is needed.
4.3.1.2 Re-tune the gains and offsets if any of the PFTBA ion
amu values specified in Section 4.3.1 are off by more than 0.2
amu.
4.3.1.3 The "tilt" of the tune is changed by
adjusting the repeller, ion focus and entrance lens voltages.
This adjustment is performed when the intensities of the 219
and/or 502 ions are not within the recommended criteria specified
in Section 4.3.1.
4.3.1.4 Clean the source when the electron multiplier voltage
must be increased to near 3000 V or tune criteria cannot be met.
After cleaning the source the electron multiplier should be set
to 1500 V or less.
4.3.1.5 IMPORTANT: Whenever any of the tune parameters are
changed it will be necessary to perform a new multi-level calibration
and to verify the lower detection limit by analyzing the 0.1 g/L
standard.
4.4 Check the system tune using 1.0 uL of 25mg/mL 4-bromofluorobenzene
(BFB) in methanol every 6 months, after cleaning the source, or
after retuning.
4.4.1. The BFB tune parameters should meet the following EPA
criteria (intensity counts are not specified by EPA):
m/z amu
|
Ion Abundance Criteria
|
50
|
15 - 40 % of 95 amu
|
75
|
30 - 60 % of 95 amu
|
95
|
Base peak 100 %
|
96
|
5 - 9 % of 95 amu
|
17
|
< 2 % of 174 amu
|
17
|
> 50 % of 95 amu
|
17
|
5 - 9 % of 174 amu
|
17
|
95 - 101 % of 174 amu
|
17
|
5 - 9 % of 176 amu
|
If the BFB criteria cannot be met, then re-tune the mass spectrometer
by the following:
4.4.1.1 If the low intensity ions (96, 175 and 177) are not
found, increase the electron multiplier voltage.
4.4.1.2 If the "tilt" is wrong (50, 75, 174 and 176
are too high or too low), then change the ion focus and entrance
lens.
4.5 Preparation of Standards
4.5.1 Prepare four different 100 g/L standards (standards according
to the groups in Section 3.3) in isooctane solvent. Prepare this
standard at least every three months.
4.5.1.1 Starting with least volatile component, add 1.00 g
of each component to a tared 10 mL volumetric flask and weigh
to 0.1 mg.
4.5.1.2 Bring to mark with isooctane.
4.5.2 Prepare 10 g/L standard by adding 1.0 mL of each of the
four 100 g/L standards to a 10 mL volumetric flask and diluting
to the mark with isooctane.
4.5.3 Prepare the 1 g/L standard by adding 1.0 mL of the 10
g/L standard to a 10 mL volumetric flask and diluting to the mark
with isooctane.
4.5.4 Prepare a 0.1 g/L standard by adding 1.0 mL of the 1 g/L
standard to a 10 mL volumetric flask and diluting to mark with
isooctane. This solution is used in Section 4.3.1.5 to verify
the lower detection limit.
4.5.5 Prepare the 25 g/L standard by combining 1.0 mL each of
the four 100 g/L standards in a 5 mL vial with a TeflonR faced
septum cap. Prepare this standard once a month.
4.6 Calibration
4.6.1 Inject a solvent blank to verify that the syringe and
the system is clean.
4.6.2 Inject separately 1.0 uL of the 1 g/L standard, the 10
g/L, the 25 g/L and each of the four 100 g/L standards using solvent
flush injection technique.
Solvent flush injection technique is described in Hewlett Packard
GC Manual (HP 5890 Series II Reference Manual 6 - 22).
4.6.2.1 For each compound, create a multilevel calibration
curve of counts vs. ug injected.
4.6.2.2 Calculate the response factor for each component.
The response factor is defined as the counts divided by the ug
injected.
4.6.2.3 Calculate the percent relative standard deviation (%RSD)
of the average response factor of each component in the multi-level
calibration. The values should be less than 25%.
4.6.3 Once a valid multi-level calibration exists, the GC/MS
is checked by comparing the response factor of the 25g/L standard
with those obtained from the valid multi-level calibration.
4.6.3.1 The percent difference (% Diff.) from the average response
factor should be less than 25% for each component.
4.6.3.1.1 If the % Diff. is greater than 25% a recalibration
is necessary if the compounds found to be out of calibration
are present in the sample.
4.6.3.2 Either prepare new standards (because standards have
degraded over time) or prepare standards which are closer to the
expected concentrations in the sample to create new calibration
reponse factors.
4.6.4 For liquid samples, the linear range is 1 to 100 ug/uL,
and the lower detection limit is 0.1 ug/uL.
4.6.5 For gas injections, a 10 ppm lower limit of detection
has been demonstrated for BTX gases.
4.6.5.1 Samples that contain less than 10 ppm must be concentrated
before being injected onto the column.
4.7 Analysis
4.7.1 Analysis of headspace gas
4.7.1.1 Inject 1 mL of ambient air taken from the vicinity
of the GC/MS. There should only be a large air peak (18, 28,
32, and 40 amu). If other peaks are present, identify and quantitate
each peak attributed to VOC. Do not quantitate the air peak.
4.7.1.2 If sample is in a container with a septum, take 1 mL
aliquot of headspace through the septum using a syringe specified
in Section 2.3.1. Do not allow the syringe tip to touch the surface
of the liquid or solid portion of the sample. Close syringe valve
before removing syringe assembly from the septum.
4.7.1.3 If sample is in a can, open lid slightly, take a 1mL
aliquot of the headspace with a syringe specified in Section 2.3.1.
Be careful not to touch the rag or the surface of the liquid
while removing headspace gas. Close syringe valve before removing
syringe assembly from the container. Immediately reclose the
container.
4.7.1.4 Inject 1 mL headspace onto the GC/MS and identify and
quantitate all peaks seen.
4.7.1.5 Subtract any VOC due to the ambient air.
4.7.2 Analysis of a liquid sample
4.7.2.1 With a syringe specified in Section 2.3.2, inject 1.0
uL isooctane using isooctane as solvent flush or other appropriate
solvent if isooctane is known to interfere. Identify and quantitate
all peaks except the solvent peak.
4.7.2.2 With a syringe specified in Section 2.3.2, inject 1.0
uL of sample using isooctane as solvent flush or other appropriate
solvent if isooctane is known to interfere. Identify and quantitate
all peaks except the solvent peak.
4.7.2.3 If any components
are over the calibration range for that component (>100 g/L),
dilute as needed.
4.7.2.4 If any components are observed but are not in the standard
(See Section 3.2 ), prepare additional standards and calibrate.
4.7.3 If components are present which cannot be identified by
matching to known spectra, provide spectra for possible future
identification. Estimate the concentration of unknown peaks by
area percent quantitation based on the Total Ion Chromatogram.
[Return to Top]
5. Calculations
5.1 Calculations
5.1.1 For headspace gas calculations, concentration of each
compound in ppm by volume:
Concentration, ppm (vol) = [24.5(A)*(103)]/[(B)*(MW)]
where:
|
A = quantified amount of a compound, ug
|
B = injection volume, mL
|
MW = molecular weight of the compound, g/mole
|
Note:
|
24.5 L/mole is the molar volume of an ideal gas at 25°C
|
5.1.2 For liquid sample calculations, concentration of each
compound in percent by volume is:
Concentration, % by vol = [(Y)*(0.1)]/[(Z)*(D)]
where:
|
Y = quantified amount of a compound for one injection, ug
|
Z = injection volume, uL
|
D = density of the compound, g/mL
|
5.1.3 For liquid samples,
concentration for each compound in g/L:
Concentration, g/L = Y/Z
where:
|
Y = quantified amount of a compound for one injection, ug
|
Z = injection volume, uL
|
5.1.3.1 Total VOC, g/L is the sum of the concentrations for
each compound.
[Return to Top]